Publisher: Invest Ophthalmol Vis Sci. 2018; 59(4), AMD152-AMD159

Authors: Susanne G. Welker; Maximilian Pfau; Manuel Heinemann; Steffen Schmitz-Valckenberg; Frank G. Holz; Robert P. Finger

Retest Reliability of Mesopic and Dark-Adapted Microperimetry in Patients With Intermediate Age-Related Macular Degeneration and Age-Matched Controls.

Abstract

Purpose: To determine the intrasession test-retest reliability of mesopic and dark-adapted fundus-controlled perimetry in patients with intermediate age-related macular degeneration (iAMD).

Methods: We conducted a cross-sectional study with 23 iAMD patients (67.3 ± 8.2 years; range, 50-85; 78% female) and 24 healthy controls (61.3 ± 5.2 years; range, 50-71; 50% female) using a modified MAIA microperimeter. All patients underwent duplicate mesopic (achromatic stimuli, 400-800 nm) and dark-adapted (red stimuli, 627 nm) microperimetry, using a grid of 33 stimuli over 14° of the central retina. Main outcome measure was the intrasession test-retest reliability for pointwise sensitivity (PWS).

Results: PWS test-retest reliability was good among mesopic and dark-adapted testing in both patients and controls (coefficient of repeatability of 4.4, 4.52, 3.96, and 4.56 dB, respectively). Mean mesopic sensitivity in patients was 2.62 dB lower than in controls (P < 0.01); mean dark-adapted sensitivity was 2.49 dB lower than in controls (P < 0.01).

Conclusions: The modified MAIA device allows for reliable mesopic and dark-adapted microperimetry in iAMD patients. We found that iAMD is associated with both reduced mesopic and dark-adapted retinal sensitivity